TEKTRONIX 2215A

Contact us for quote

Click to download this page as a PDF-Document. Download this page as a PDF-Document.
Product Details

60 MHz Oscilloscope

 2mV SENSIVITY, 5nS SWEEP RATE, VARIABLE SWEEP DELAY, WITH AN IMPROVED CRT INTENSITY OVER THE PLAIN MODEL.

Product Specifications

VERTICAL DEFLECTION (2 identical channels)
Bandwidth*1 and Risetime*20°C to +40°C+40°C to +50°C
Dc to 60 MHz, 20 mV/div to 10 V/div, 5.8 ns reduced
Dc to 50 MHz, 2 mV to 10 mV/div, 7 ns
Dc to 50 MHz, 2 mV to 10 mV/div 7 ns
*1 Measured at -3 dB.
*2 At all deflection factors from 50Ω  terminated source.
Deflection Factor2 mV/div to 10 V/div ± 3% (+20°C to +30°C) or ± 4% (0°C to +50°C). 1-2-5 sequence.  Uncalibrated, continuously variable between steps to at least 25 V/div.
Display ModesCH 1, CH 2, CH 2 Add (normal and inverted), Alternate, Chopped: approx. 250 kHz rate, electronically switched.
Common-Mode Rejection RatioAt least 10:1 at 10 MHz for common-mode signals of 6 divisions or less.
Input R and C1 MΩ  ± 2% paralleled by approx. 30 pF.
Maximum Input VoltageDc Coupled400 V (dc + peak ac)
800 V (p-p ac at 1 kHz or less)
Ac Coupled400 V (dc + peak ac)
800 V (p-p ac at 1 kHz or less)
Delay LinePermits viewing leading edge of displayed waveform
HORIZONTAL DEFLECTION
Time Base A (2213 and 2215)0.05 µs/div to 0.5 s/div (1-2-5 sequence).  10X magnifier extends max sweep rate to 5 ns/div.
Time Base B (2215 only)0.05 µs/div to 50 ms/div (1-2-5 sequence). 10X magnifier extends max sweep rate to 5 ns/div.
Variable Time Control (2213 and 2215)Time Base A provides continuously variable uncalibrated sweep rates between steps to at least 1.25 s/div.
Time Base A (2213 and 2215) and B (2215 only) Accuracy*1
 +20°C to +30°C0°C to + 50°C
Unmagnified± 3%±4%
Magnified± 5%±6%
*1 Center 8 divisions.
Horizontal Display Modes (2213)A, A intensified after delay, delayed.
Horizontal Display Modes (2215)A, alternate (A intensified by B and B), B. Electronic switching between intensified and delayed sweep.
2213 SWEEP DELAY
Delay Times< 0.5 µs, 10 µs, and 0.2 ms.
MultiplierIncreases delay time by 20 to 1 or more.
Jitter5000 to 1 (0.02%) of maximum available delay time.
2215 SWEEP DELAY
Delay TimesContinuously variable by means of a 10 to 1 vernier control.  Delayed (B) portion is intensified on the main (A) trade.
Delay Position Range< 0.5 div + 300 ns to more than 10 div.
Delay Dial Accuracy± 1.5% of reading past 1 div.
A/B Sweep SeparationControl permits main and delayed sweep to be separated by at least 3.5 div.
Jitter10,000 to 1 (0.01%) of maximum available delay time.
TRIGGERING
2213 and 2215 A Time Base Trigger ModesNormal (sweep runs when triggered), automatic (sweep runs in the absence of a triggering signal and triggers automatically for signals down to 20 Hz), and TV field (with slope set for negative going transitions, and trigger level adjusted close to blanking level, sweep starts at first line of video; use Normal for TV line display). LED indicates when sweep is triggered.
A Trigger HoldoffAdjustable control permits a stable presentation of repetitive complex waveforms.
SensitivityAuto and Normal Internal: Below 2 MHz, signal must be at least 0.4 div amplitude; requirements increase above 2 MHz; at 60 MHz, signal must be at least 1.5 div amplitude.
Auto and Normal ExternalUp to 2 MHz, trigger signal must be at least 50 mV p-p; requirements increase up to 60 MHz, where signal must be at least 250 mV p-p.
TV FieldComposite video must be at least 2 div amplitude.
Level and Slope (Norm Mode)Internal: Trigger level can be adjusted over the range of amplitudes displayed on the CRT.  
External, Dc Coupled: Level can be adjusted over a range of at least ± 2 V, or 4 V p-p.  
External; Dc Coupled and Attenuated (divided by 10): Level can be adjusted over a range of at least ± 20 V, or 40 V p-p.
External InputsR and C approx. 1 MΩ  paralleled by approx. 30 pF.
2215 Delayed (B) Timebase
Level and SlopeSeparate slope and level controls for triggering B sweep.
SensitivityUp to 2 MHz, signal must be at least 0.4 div in vertical amplitude; requirements increase up to 60 MHz, where signal must be at least 2 div in amplitude.
X-Y OPERATION
Full Sensitivity X-Y (CH1 Horizontal, CH 2 Vertical)2 mV/div to 10 V/div accurate ± 5%.  Bandwidth is dc to at least 2 MHz.  Phase difference between amplifiers is 3° or less from dc to 50 kHz.
DISPLAY
CRT8 X 10 cm display.   Horizontal and vertical center lines further marked in 0.2 cm increments.  GH (P31) Phosphor standard.  10 kV accelerating potential, mesh grid, halo suppressed.
GraticuleInternal, non-parallax, not illuminated; markings for measurement of risetime.
Beam FinderCompresses trace to within graticule area for ease in locating an off-screen signal.  A preset intensity level provides a constant brightness.
Z-Axis InputDc coupled, positive-going signal decreases intensity; 5 V p-p signal causes noticeable modulation at normal intensity; dc to 5 MHz.
ENVIRONMENTAL CHARACTERISTICS
Ambient TemperatureOperating: 0°C to +50°C.   Nonoperating: -55°C to +75°C.
AltitudeOperating: To 4600 m (15,000 ft); maximum allowable ambient temperature decreased by 1°C/1000 ft from 5000 to 15,000 ft. Nonoperating:  15,000 m (50,000 ft).
VibrationOperating test samples were subjected to sinusoidal vibration in the X, Y and Z axis with the frequency varied from 10 Hz to 55 Hz to 10 Hz in 1 minute sweeps for a duration of 15 minutes per axis and a dwell of 10 minutes at 55 Hz.  Total displacement was 0.015 in p-p (2.4 g's at 55 Hz).
HumidityOperating and Nonoperating: Test samples were subjected to 5 cycles (120 hours) of humidity testing.
ShockOperating and Nonoperating: Test samples were subjected to 3 shocks, both directions along each axis for a total of 18 shocks.  Peak accelerations of each 1/2-sine shock were 30 g's.
OTHER CHARACTERISTICS
Probe Adjust SignalSquarewave, 0.5 V ± 20%, 1 kHz ± 20%.
Power Requirements90 to 250 V, 48 to 440 Hz without range switching, 50 Ω  maximum at 115 V and 60 Hz.
PHYSICAL CHARACTERISTICS
Cabinet Dimensionsmmin
Width (with handle)
Width (without handle)
Height (with feet and handle)
Depth (with front cover)
Depth (without front cover)
Depth (with handle extended)
360
237
137
445
440
511
14.2
12.9
5.4
17.5
17.3
20.1
Weights approx.kglb
Net (with cover accessories and pouch)
Net (without cover accessories and pouch)
Shipping (domestic)
7.6

6.1

8.2

16.8

13.5

18.0

INCLUDED ACCESSORIESTwo P6122 10X voltage probes (010-6122-01); operator's manual, service manual