TEKTRONIX A6303 CURRENT PROBE

Contact us for quote

Click to download this page as a PDF-Document. Download this page as a PDF-Document.
Product Details

CURRENT PROBE

The larger A6303 measures higher amplitude, lower frequency currents. It clips onto conductors up to 21.1 mm (0.83 in.) in diameter. Its frequency range is DC to 15 MHz. It will measure continuous currents up to 100 A and peak currents up to 500 A, as long as the Amp*second product of 10,000 A*μs is not exceeded.

  • 100 A Ac and Dc Current Measurements
  • Peak Pulse Measurement to 500 A
  • Ac or Dc Coupling
  • 25 x 21 mm (1 x 0.83 in) Jaw Opening

A6303 with AM503 (Current Probe Amplifier)
Sensitivity: Scope @ 10 mV/div10 mA/div to 50 A/div
BandwidthDc to 15 MHz
Rise Time23 ns
Max Ac Current CW200 A p-p
Derated Above20 kHz  (12 A @ 10 MHz)
Maximum Current
(Peak Not To Exceed A-S Product)
500 A
A-S Product10,000 x 10-6
Insertion Z0.02 ohms @ 1 MHz
0.15 ohms @ 15 MHz
Max Hardware Volts700 V
Max Conductor Diameter0.83 inch
Cable Length2 m
Magnetic Susceptibility25 mA/Gauss